Alarms are sound from the ventilator indicates that there is a problem that must be solved so as not to harm the patient. There are several types of mechanical ventilator alarms that are designed to detect different issues. Some of the most common alarms include:

  • Ventilator Inoperative
The Problem It means that the ventilator is not working due to a technical malfunction.
The SolutionDisconnect the patient from the ventilator and work in an AMBU bag with 100% oxygen until the patient is connected to another ventilator and the doctor is notified to examine the patient and inform the technician to check the device for malfunction.
  • High Pressure
The ProblemIt means that the pressure inside the ventilator circuit is too high. This could be caused by a blockage in the breathing tube, a patient coughing or gagging. Other causes may include;
1.Increases secretion in the airways and the tube.
2.One of the ventilator connections bended.
3.The patient is biting the tube.
4.Finding the endotracheal tube has moved and entered a little towards the right and you can make sure by hearing the patient’s chest with a stethoscope and noticing that you hear the sound of air in one side.
The Solution1.Suction to remove the secretions. 2.Examine all connections of the ventilator to find out the problem and solve it, whether by changing the connection or modifying it.
3.The patient is placed on the oral airway in his mouth and it is fixed tightly so that it does not come out of his mouth, or the patient is given the dose of the dormicam that he is supposed to take.
4.Change the endotracheal tube position to it’s normal one.
  • Low Pressure
The Problem It means that the pressure inside the ventilator circuit is too low. Causes may include;
1.A hole in any of the ventilator connections.
2.Disconnecting of any of the ventilator connections or may be a disconnecting in endotracheal tube connection that connect the ventilator to the patient (Y).
3.A leakage from the humidifier.
4.A leakage from the endotracheal tube cuff and this cuff is important for fixing the tube and thus preventing the air leakage.
The Solution 1.Change the connection.
2.Examin all the connections and making sure that they are connected.
3.Change the humidifier and inform the technician.
4.Change the endotracheal tube.
  • High Respiratory Rate
The Problem It means that the patient’s respiratory rate is too high. This could be caused by a variety of factors, such as anxiety, pain, or a fever.
The SolutionInform the doctor to write an order because the treatment is different from case to case.
  • Apnea
The Problem It means that the patient doesn’t take any breath on his own.
The SolutionEncourage the patient to take the breath or giving the patient one breath with the AMBU bag or changing the mode.
  • Low Tidal Volume
The Problem It means that the volume of air being delivered to the patient’s lungs is too low. This could be caused by a variety of factors, such as a leak in the breathing tube.
The SolutionInform the technician and change the oxygen source.

How Mechanical Ventilator Alarms Work

Mechanical ventilator alarms are designed to alert clinicians to potential problems with the device or the patient. When an alarm is triggered, it will sound an audible alarm and display a message on the machine’s screen. The clinician can then investigate the cause of the alarm and take appropriate action to correct the problem.

In some cases, the alarm may be triggered by a transient issue, such as a cough or a brief drop in pressure. In these cases, the clinician may simply need to adjust the ventilator settings or reassure the patient. However, in other cases, the alarm may be indicative of a serious problem, such as a disconnection in the breathing tube or a malfunctioning ventilator. In these cases, the clinician must take immediate action to correct the problem, which may include repositioning the patient, adjusting the ventilator settings, or replacing a faulty component.

What to Do When a Mechanical Ventilator Alarm is Triggered

When a mechanical ventilator alarm is triggered, the clinician should first assess the patient to determine if they are in distress. If the patient appears stable, the clinician can investigate the cause of the alarm by checking the breathing tube for disconnections or leaks, adjusting the ventilator settings, or performing other troubleshooting steps.

If the patient appears to be in distress, the clinician should take immediate action to correct the problem. This may include manually ventilating the patient, repositioning the patient, or replacing a faulty component. In some cases, the patient may need to be transferred to a different ventilator or to another type of respiratory support.